Invacare Camping Equipment CG8000 User Manual

Invacare 8000 Alternating Pressure Mattress  
Assembly, InstallationandOperatingInstructions  
Model No. CG8000  
NOTE: The above model number contains one (1) Alternating Pressure Mattress and one (1)  
Power Unit.  
NOTE: Check ALL parts for shipping damage. In case of shipping damage, DO NOT use. Contact  
Carrier/Dealer for further instruction.  
SAFETY SUMMARY  
To ensure the safe installation of the Invacare 8000 Alternating Pressure  
Mattress and Power Unit, these instructions MUST be followed:  
WARNING  
DO NOT install or use this equipment without first reading and  
understanding this instruction sheet. If you are unable to understand the  
Warnings, Cautions or Instructions, contact a healthcare professional,  
dealer or technical personnel before attempting to install this equipment -  
otherwise, injury or damage may occur.  
DO NOT use the mattress near an open flame or while smoking.  
Inspect power cord, plug and air hoses of the Invacare 8000 Alternating  
Pressure Mattress/Power Unit for breaks, tears, or kinks. If damage is  
found, DO NOT use.  
DO NOT, under any circumstances, cut or remove the round grounding  
prong from any plug used with or for Invacare products. Devices are  
equipped with three-prong (grounding) plugs for protection against  
possible shock hazards. Where a two-prong wall receptacle is encountered,  
it is the personal responsibility and obligation of the customer to contact a  
qualified electrician and have the two-prong receptacle replaced with a  
properly grounded three-prong wall receptacle in accordance with the  
National Electrical Code. If you must use an extension cord, use ONLY a  
three-wire extension cord having the same or higher electrical rating as the  
device being connected.  
DO NOT place hot objects (for example: heating pads, electric blankets,  
etc.) on top of the mattress. This may weaken the mattress material.  
Use extreme care when handling sharp objects such as needles near the  
mattress.  
Do NOT inflate mattress of an OCCUPIED bed.  
The Alternating Pressure Mattress has a weight limitation of 250 lbs. (114 kg.).  
The Invacare 8000 APM MUST be installed on medical bed frames or medical-  
grade home care beds with side rails. After placing the patient on the mattress,  
verify the side rails provide adequate protection for the patient. The side rails  
MUST be in the raised position, whenever a patient is on the bed.  
INVACARE 8000 PRESSURE MATTRESS  
Part No. 1078996 Rev. B  
1
 
Power Unit Air  
Hose Quick  
Disconnect  
Power  
Unit  
Cushion  
Fittings  
Transport Mat  
Cushion  
Mattress Air Hose  
Quick Disconnect  
Power  
Unit  
Knob  
CPR Plugs  
+MAX  
Setting  
Power  
Switch  
-MIN  
Setting  
FIGURE 2- INFLATING THE MATTRESS  
INSTALLING/REMOVING THERAPY PAD (FIGURE 3)  
INSTALLING  
1. Position therapy pad over mattress.  
2. Insert one (1) of the ten (10) straps into  
open latch as shown in FIGURE 3.  
3. Pull strap until therapy pad fits tight  
around mattress.  
4. Push down on latch to secure.  
Strap  
5. Repeat STEPS 2 - 4 for remaining  
Latch  
nine (9) straps.  
6. Secure mattress base to the bed frame.  
Refer to SECURING MATTRESS  
BASE TO BED FRAME in this  
instruction sheet.  
REMOVING  
1. Remove patient from bed.  
2. Pull up on one (1) of the ten (10)  
FIGURE 3 - INSTALLING/REMOVING  
THERAPY PAD  
latches that secure the therapy pad  
to the mattress.  
3. Remove strap from latch.  
4. Repeat STEPS 2 - 3 for remaining nine (9) latches.  
5. Remove therapy pad from mattress.  
Part No. 1078996 Rev. B  
INVACARE 8000 PRESSURE MATTRESS  
3
 
SECURING MATTRESS BASE TO BED FRAME (FIGURE 4)  
1. Wrap one (1) of the ten (10) straps on  
the mattress around the bed frame.  
Double D-rings  
2. Insert strap into the two (2) double D-  
rings.  
3. Feed strap back through only one (1) of  
the double D-rings.  
4. Pull the strap tight to secure the mattress  
base to the bed frame.  
5. Repeat STEPS 1 - 4 for the remaining  
nine (9) double D-rings.  
NOTE: Each SIDE contains three (3) double D-rings  
and each END contains two (2) double D-rings for  
a total of ten (10) double D-rings.  
6. Adjust the mattress to the proper  
firmness. Refer to PLACING  
PATIENT ON MATTRESS AND  
ADJUSTING FIRMNESS in this  
instruction sheet.  
FIGURE 4 - SECURING MATTRESS BASE  
TO BED FRAME  
PLACING PATIENT ON MATTRESS AND ADJUSTING  
FIRMNESS (FIGURE 5)  
NOTE: Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet.  
1. Place patient on the mattress.  
2. Check the level that the patient sinks into mattress.  
NOTE: The pressure of the power unit can be adjusted to provide maximum effectiveness and  
pressure relief for the patient.  
NOTE: The pressure can be adjusted with a range from -MIN to +MAX.  
NOTE: Every other cushion inside the mattress is connected together known as a chamber. The  
mattress consists of two (2) chambers.  
NOTE: The power unit will cycle every eight (8) minutes. It will alternate inflating and deflating the  
two (2) air chambers.  
NOTE: The patients body should sink approximately 1/3 into the mattress.  
3. If the level is unacceptable, do one (1) of the following:  
A. To increase firmness - turn knob towards +MAX setting.  
B. To decrease firmness - turn knob towards -MIN setting.  
NOTE: After adjusting the knob, wait ten (10) minutes before rechecking the level the patient sinks  
into mattress.  
4. Repeat STEP 3 until the patients body sinks 1/3 into the mattress.  
INVACARE 8000 PRESSURE MATTRESS  
Part No. 1078996 Rev. B  
4
 
Power Unit  
Knob  
Power Switch  
+MAX Setting  
-MIN  
Setting  
FIGURE 5 - PLACING PATIENT ON MATTRESS  
AND ADJUSTING FIRMNESS  
DEFLATING MATTRESS (FIGURE 6)  
NOTE: Alternating cushions are connected together to form a chamber. The mattress consists of  
two (2) chambers.  
DETAIL “A”  
STANDARD DEFLATION  
Power Unit  
1. Remove therapy pad. Refer to  
Power Unit Air Hose  
Quick Disconnect  
INSTALLING/REMOVING  
THERAPY PAD in this instruction sheet.  
2. Disconnect the two (2) power unit air  
hose quick disconnects from the two (2)  
mattress air hose quick disconnects as  
shown in DETAIL “A”.  
Mattress Air  
Hose Quick  
Disconnect  
3. Fold the mattress in half to force the air out.  
4. Before re-inflation is possible, connect  
the two (2) power unit air hose quick  
disconnects to the two (2) mattress air  
hose quick disconnects.  
Manifold  
5. Inflate the mattress. Refer to  
INFLATING THE MATTRESS in  
this instruction sheet.  
RAPID (CPR) DEFLATION  
CPR Plugs  
1. Pull the two (2) CPR straps labeled  
“PULL CPR” away from mattress.  
NOTE: Using the “PULL CPR” straps will disconnect  
the CPR plugs from the manifold, which allows the  
two (2) chambers to deflate.  
2. Before re-inflation is possible, install the  
CPR plugs on the CPR straps into the  
fittings on the manifold.  
3. Inflate the mattress. Refer to  
INFLATING THE MATTRESS in  
this instruction sheet.  
FIGURE 6 - DEFLATING MATTRESS  
Part No. 1078996 Rev. B  
INVACARE 8000 PRESSURE MATTRESS  
5
 
REPLACING CUSHION(S) (FIGURE 7)  
NOTE: Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet.  
1. Remove patient from bed.  
2. Remove the therapy pad. Refer to INSTALLING/REMOVING THERAPY PAD  
in this instruction sheet.  
3. Unbutton the EXISTING cushion from the mattress base.  
4. Repeat STEP 3 for opposite side of the cushion.  
5. Pull the cushion fitting away from the manifold fitting to remove the cushion from  
the mattress.  
NOTE: Alternating cushions are connected together to form a chamber. The mattress consists of  
two (2) chambers.  
NOTE: The cushions in the same chamber as the one removed in STEP 5 will deflate when the  
fittings are removed.  
6. Position NEW cushion into mattress.  
7. Insert cushion fitting into manifold fitting to secure.  
8. Button the cushion(s) to secure to the mattress base.  
9. If more cushions need replacing, repeat STEPS 3 - 8.  
10. Inflate the mattress. Refer to INFLATING THE MATTRESS in this instruction  
sheet.  
Therapy  
Pad  
Button Holes  
Manifold  
Manifold Fitting  
Mattress Base  
Cushion Fitting  
Button  
Cushion  
FIGURE 7- REPLACING CUSHION(S)  
CARE AND MAINTENANCE  
1. The mattress can be cleaned by wiping it with a germicidal disinfectant solution.  
2. The mattress therapy pad is machine washable. Air dry at a cool temperature.  
INVACARE 8000 PRESSURE MATTRESS  
Part No. 1078996 Rev. B  
6
 
LIMITED WARRANTY  
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY  
WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED  
AFTER JULY 4, 1975.  
This warranty is extended only to the original purchaser/user of our products.  
This warranty gives you specific legal rights and you may also have other legal rights  
which vary from state to state.  
Invacare warrants its product to be free from defects in materials and workmanship  
for one(1) year on the pump and six (6) months on the mattress for the original  
purchaser. If within such warranty period any such product shall be proven to be  
defective, such product shall be repaired or replaced, at Invacare's option. This  
warranty does not include any labor or shipping charges incurred in replacement part  
installation or repair of any such product. Invacare's sole obligation and your exclusive  
remedy under this warranty shall be limited to such repair and/or replacement.  
For warranty service, please contact the dealer from whom you purchased your  
Invacare product. In the event you do not receive satisfactory warranty service, please  
write directly to Invacare at the address on the next page. Provide dealer's name,  
address, model number, date of purchase, indicate nature of the defect and, if the  
product is serialized, indicate the serial number.  
Invacare Corporation will issue a return authorization. The defective unit or parts  
must be returned for warranty inspection using the serial number, when applicable, as  
identification within thirty (30) days of return authorization date. DO NOT return  
products to our factory without our prior consent. C.O.D. shipments will be refused;  
please prepay shipping charges.  
LIMITATIONS AND EXCLUSIONS: THIS WARRANTY SHALL NOT APPLY TO  
PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO  
THE ENCLOSED INSTRUCTIONS. IN ADDITION, THE FOREGOING  
WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE  
SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS  
SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION,  
MAINTENANCE OR STORAGE; PRODUCTS MODIFIED WITHOUT  
INVACARE'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT  
LIMITED TO: MODIFICATION THROUGH THE USE OF UNAUTHORIZED  
PARTS OR ATTACHMENTS): PRODUCTS DAMAGED BY REASON OF  
REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT  
OF INVACARE; PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND  
INVACARE'S CONTROL; PRODUCTS REPAIRED BY ANYONE OTHER THAN  
AN AUTHORIZED INVACARE DEALER. SUCH EVALUATION SHALL BE  
SOLELY DETERMINED BY INVACARE.  
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER  
WARRANTIES WHATSOEVER, WHETHER EXPRESSED OR IMPLIED,  
INCLUDING THE IMPLIED WARRANTIES OF MERCHANTIBILITY AND  
FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR  
VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO  
REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PERSUANT TO  
THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED  
WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION  
OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT  
BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES  
WHATSOEVER.  
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/  
PROVINCIAL LAWS AND REQUIREMENTS.  
Part No. 1078996 Rev. B  
INVACARE 8000 PRESSURE MATTRESS  
7
 
Invacare Corporation www.invacare.com  
USA  
Canada  
One Invacare Way  
Elyria, Ohio USA  
44036-2125  
570 Matheson Blvd E Unit 8 Invacare is a registered trademark of  
Mississauga Ontario  
L4Z 4G4 Canada  
800-668-5324  
Invacare Corporation.  
Yes, you can. is a trademark of Invacare  
Corporation.  
800-333-6900  
© 2003 Invacare Corporation  
Part No. 1078996  
Rev B - 05/03  
 

Indesit Oven FIMU 23 BK User Manual
InFocus Home Theater System IN118HDa User Manual
Intel Computer Hardware 631xESB User Manual
Iomega Server ix4 200r User Manual
IRiver MP3 Player IHP 140 User Manual
John Deere Lawn Mower 5045E User Manual
Keys Fitness Home Gym ST MAB User Manual
King Canada Grinder KC 690L User Manual
Kodak Digital Camera C1013 User Manual
Kodak Digital Camera CX4210 User Manual